Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 8062A: (1 year) CAL VER /5520 DATE: 20-Aug-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 48 NUMBER OF LINES: 145 CONFIGURATION: Fluke 5520A ============================================================================= # # Source: # Fluke 8062A instruction manual (P/N 632695, March 1982, Rev. 2, 3/93) # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5520A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- N P F W 1.002 ASK+ K 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Turn on the UUT. 1.004 DISP 1.004 DISP Before performing calibration allow the UUT to stabilize 1.004 DISP (30 min.) @ 23+/-5 degC. Humidity less than 80%. 1.005 HEAD {MICROCOMPUTER AND DISPLAY TEST} 1.006 DISP Use the automatic power-on self-test to test the 1.006 DISP microcomputer and the LCD display. Turn the UUT 1.006 DISP off, then turn it on while observing the display. 1.006 DISP All of the LCD segments should turn on. After one 1.006 DISP or two seconds, the display should go blank briefly 1.006 DISP and then respond to switch selections. 1.007 EVAL Did the {display test} pass? 2.001 HEAD {DC VOLTAGE} 2.002 DISP Connect the 5520A to the UUT as follows: 2.002 DISP [32] 5520A NORMAL HI to UUT V/Ohm 2.002 DISP [32] 5520A NORMAL LO to UUT COMMON 2.003 HEAD DC VOLTAGE: {200mV Range} 2.004 5520 200 190.00mV 0.11U 2W 3.001 5520 200 -190.00mV 0.11U 2W 4.001 HEAD DC VOLTAGE: {2V Range} 4.002 5520 2 1.9000V 0.0011U 2W 5.001 HEAD DC VOLTAGE: {20V Range} 5.002 5520 20 19.000V 0.015U 2W 6.001 HEAD DC VOLTAGE: {200V Range} 6.002 5520 200 190.00V 0.15U 2W 7.001 HEAD DC VOLTAGE: {1000V Range} 7.002 5520 1000 1000.0V 0.9U 2W 8.001 HEAD {AC VOLTAGE} 8.002 HEAD AC VOLTAGE: {200mV Range} 8.003 5520 200 100.00mV 0.60U 200H SI 2W 9.001 5520 200 100.00mV 1.40U 20kH SI 2W 10.001 HEAD AC VOLTAGE: {2V Range} 10.002 5520 2 1.0000V 0.0060U 200H SI 2W 11.001 5520 2 1.0000V 0.0070U 1kH SI 2W 12.001 5520 2 1.0000V 0.0070U 10kH SI 2W 13.001 5520 2 1.0000V 0.0140U 30kH SI 2W 14.001 5520 2 1.0000V 0.0110U 20H SI 2W 15.001 5520 2 0.1000V 0.0015U 200H SI 2W 16.001 5520 2 0.1000V 0.0050U 30kH SI 2W 17.001 HEAD AC VOLTAGE: {20V Range} 17.002 5520 20 10.000V 0.060U 200H SI 2W 18.001 5520 20 10.000V 0.520U 10kH SI 2W 19.001 5520 20 10.000V 0.540U 30kH SI 2W 20.001 HEAD AC VOLTAGE: {200V Range} 20.002 5520 200 100.00V 0.60U 200H SI 2W 21.001 5520 200 100.00V 5.20U 20kH SI 2W 22.001 HEAD AC VOLTAGE: {750V Range} 22.002 5520 750 750.0V 16.0U 100H SI 2W 23.001 5520 750 750.0V 16.0U 1kH SI 2W 24.001 HEAD {RESISTANCE} 24.002 DISP Connect the 5520A to the UUT as follows: 24.002 DISP [32] 5520A NORMAL HI to UUT V/Ohm 24.002 DISP [32] 5520A NORMAL LO to UUT COMMON 24.002 DISP [32] 5520A AUX HI to UUT V/Ohm 24.002 DISP [32] 5520A AUX LO to UUT COMMON 24.003 HEAD RESISTANCE: {200 Ohm Range} 24.004 5520 200 0.00Z 0.04U CW 25.001 5520 200 100.00Z 0.14U CW 26.001 HEAD RESISTANCE: {2 kOhm Range} 26.002 5520 2 1.0000kZ 0.0012U CW 27.001 HEAD RESISTANCE: {20 kOhm Range} 27.002 5520 20 10.000kZ 0.012U CW 28.001 HEAD RESISTANCE: {200 kOhm Range} 28.002 PIC 55_V8062 28.003 DISP Connect the 5520A to the UUT as follows: 28.003 DISP [32] 5520A NORMAL HI to UUT V/Ohm 28.003 DISP [32] 5520A NORMAL LO to UUT COMMON 28.004 5520 200 100.00kZ 0.12U 2W 29.001 HEAD RESISTANCE: {MOhm Range} 29.002 DISP Set UUT to the MOhm range. 29.003 5520 1.0000MZ 0.0022U 2W 30.001 5520 10.00MZ 0.05U 2W 31.001 5520 40.0MZ 0.7U 2W 32.001 5520 290MZ 8U 2W 33.001 HEAD {CONTINUITY TEST} 33.002 5520 2 0.1000kZ 0.0003U 2W 34.001 DISP Press the -><-))) button. 34.002 EVAL Is the {-><- symbol visable on the display}? 35.001 DISP Press the -><- button again to enable the audible tone. 35.002 5520 0.1000kZ S 2W 35.003 EVAL Is the {audible tone on}? 36.001 5520 * S 36.002 EVAL Is the {audible tone off}? 37.001 EVAL Is the UUT {displaying OL}? 38.001 HEAD {DIODE} 38.002 DISP Select the DIODE Function. 38.003 EVAL Is the UUT {displaying OL}? 39.001 5520 1.0000kZ 0.12U 2W 40.001 HEAD {DC CURRENT} 40.002 DISP Connect the 5520A to the UUT as follows: 40.002 DISP [32] 5520A AUX HI to UUT mA 40.002 DISP [32] 5520A AUX LO to UUT COMMON 40.003 HEAD DC CURRENT: {200uA Range} 40.004 5520 200 190.00uA 0.59U 2W 41.001 5520 200 -190.00uA 0.59U 2W 42.001 HEAD DC CURRENT: {2mA Range} 42.002 5520 2 1.9000mA 0.0059U 2W 43.001 HEAD DC CURRENT: {20mA Range} 43.002 5520 20 19.000mA 0.059U 2W 44.001 HEAD DC CURRENT: {200mA Range} 44.002 5520 200 190.00mA 1.35U 2W 45.001 HEAD DC CURRENT: {2000mA Range} 45.002 5520 2000 1900.0mA 13.5U 2W 46.001 5520 2000 -1900.0mA 13.5U 2W 47.001 HEAD {AC CURRENT} 47.002 HEAD AC CURRENT: {20mA Range} 47.003 5520 20 19.000mA 0.152U 1kH SI 2W 48.001 END